Cantilever-enhanced dual-comb photoacoustic spectroscopy
Jiapeng Wang,Hongpeng Wu,Xiaoli Liu,Gang Wang,Yong Wang,Chaofan Feng,Ruyue Cui,Zhenfeng Gong,Lei Dong
DOI: https://doi.org/10.1016/j.pacs.2024.100605
IF: 9.656
2024-08-01
Photoacoustics
Abstract:Dual-comb photoacoustic spectroscopy (DC-PAS) advances spectral measurements by offering high-sensitivity and compact size in a wavelength-independent manner. Here, we present a novel cantilever-enhanced DC-PAS scheme, employing a high-sensitivity fiber-optic acoustic sensor based on an optical cantilever and a non-resonant photoacoustic cell (PAC) featuring a flat-response characteristic. The dual comb is down-converted to the audio frequency range, and the resulting multiheterodyne sound waves from the photoacoustic effect, are mapped into the response frequency region of the optical cantilever microphone. This cantilever-enhanced DC-PAS method provides advantages such as high sensitivity, compact design, and immunity to electromagnetic interference. Through 10 seconds averaging time, the proposed approach experimentally achieved a minimum detection limit of 860 ppb for acetylene. This technology presents outstanding opportunities for highly sensitive detection of trace gases in a wavelength-independent manner, all within a compact volume.
engineering, biomedical,radiology, nuclear medicine & medical imaging,instruments & instrumentation