Ultra-high sensitive multi-pass absorption enhanced fiber-optic photoacoustic gas analyzer

Xinyu Zhao,Ke Chen,Min Guo,Chenxi Li,Chenyang Li,Guangyin Zhang,Zhenfeng Gong,Zhi Zhou,Wei Peng
DOI: https://doi.org/10.1109/tim.2022.3188043
IF: 5.6
2022-01-01
IEEE Transactions on Instrumentation and Measurement
Abstract:We proposed and designed a high-sensitivity fiber-optic photoacoustic gas analyzer based on resonant Herriott multi-pass cell (MPC) and fiber-optic cantilever acoustic sensor, which realized the double enhancement of the excitation and detection of the photoacoustic signal. The chamber volume of the miniature photoacoustic cell was 157 mL. The laser was reflected 20 times and reached an optical path length of 4 m in the PA cell, achieving the excitation enhancement of the photoacoustic signal. We conducted a comparative experiment between the Herriott-MPC and the single-pass cell, the results showed that the 2f signal of the photoacoustic system based on Herriott-MPC was enhanced by about 16 times. By using a near-infrared laser with a center wavelength of 1651 nm, the performance of the PA gas analyzer for CH4 detection was evaluated. Experimental results show that the minimum detection limit (MDL) of the system is achieved to be 3.7 ppb. Furthermore, the calculated normalized noise equivalent absorption (NNEA) coefficient is equal to 8.3×10-10 cm-1WHz-1/2. The continuous monitoring of CH4 in the environment for five consecutive days proved the feasibility and stability of the trace gas analyzer.
engineering, electrical & electronic,instruments & instrumentation