High-Sensitivity Silicon Cantilever-Enhanced Photoacoustic Spectroscopy Analyzer with Low Gas Consumption

Min Guo,Ke Chen,Chenxi Li,Lin Xu,Guangyin Zhang,Nan Wang,Chenyang Li,Fengxiang Ma,Zhenfeng Gong,Qingxu Yu
DOI: https://doi.org/10.1021/acs.analchem.1c04309
IF: 7.4
2022-01-06
Analytical Chemistry
Abstract:A silicon cantilever-enhanced photoacoustic spectroscopy (PAS)-based trace gas analyzer with low gas consumption is presented. A silicon cantilever-based fiber-optic Fabry-Perot (F-P) interferometric acoustic sensor with a compact structure and high sensitivity is designed for photoacoustic signal detection. The non-resonant photoacoustic cell (PAC) is a cylindrical copper tube with a volume of 0.56 mL. A near-infrared laser with a center wavelength of 1532.83 nm amplified using an erbium-doped fiber application amplifier is used as the excitation light. The wavelength modulation spectroscopy (WMS) technique is employed in the present work for second-harmonic photoacoustic signal detection. The experimental results show that the minimum detection limit of C<sub>2</sub>H<sub>2</sub> is 199.8 parts per trillion (ppt) with an average time of 60 s. The normalized noise equivalent absorption coefficient is calculated as 1.72 × 10<sup>-9</sup> cm<sup>-1</sup> W/Hz<sup>1/2</sup>. Furthermore, the proposed silicon cantilever-enhanced non-resonant PAS-based gas analyzer can not only analyze the gas concentration in a closed small-capacity PAC with low gas consumption but also detect target gas leakage in real time at a long distance.