High sensitivity and ultra-low concentration range photoacoustic spectroscopy based on trapezoid compound ellipsoid resonant photoacoustic cell and partial least square

Qiaoyun Wang,Shunyuan Xu,Ziheng Zhu,Jilong Wang,Xin Zou,Chu Zhang,Qiang Liu
DOI: https://doi.org/10.1016/j.pacs.2023.100583
IF: 9.656
2024-01-01
Photoacoustics
Abstract:A high sensitivity and ultra-low concentration range photoacoustic spectroscopy (PAS) gas detection system, which was based on a novel trapezoid compound ellipsoid resonant photoacoustic cell (TCER-PAC) and partial least square (PLS), was proposed to detect acetylene (C 2 H 2 ) gas. In the concentration range of 0.5 ppm ~ 10.0 ppm, the limit of detection (LOD) values of TCER-PAC-based PAS system without data processing was 66.4 ppb, which was lower than that of the traditional trapezoid compound cylindrical resonant photoacoustic cell (TCCR-PAC). The experimental results indicated that the TCER-PAC had higher sensitivity than of TCCR-PAC. Within the concentration range of 12.5 ppb ~ 125.0 ppb, the LOD and limit of quantification (LOQ) of TCER-PAC-based PAS system combined with PLS regression algorithm were 1.1 ppb and 3.7 ppb, respectively. The results showed that higher detection sensitivity and lower LOD were obtained by PAS system with TCER-PAC and PLS than that of TCCR-PAC-based PAS system.
engineering, biomedical,radiology, nuclear medicine & medical imaging,instruments & instrumentation
What problem does this paper attempt to address?