Development of Experimental Apparatus for High-Accuracy Vapor-Liquid Equilibrium(VLE) Measurement at the Temperature Ranges Of(80~180) K

Han Xiaohong,Xu Yingjie,Xu-wei Min,Gao Zanjun,Qin Yu,Guangming Chen,Qin Wang
2012-01-01
Abstract:The experimental study of VLE at low temperature is important in the chemical engineering,energy engineering,etc.,and its experimental results are always accepted by professional database.In this paper,an experimental apparatus with high accuracy is developed to get the VLE data at low temperatures.The apparatus consists of the stainless steel equilibrium cell,the temperature control system and measurement system,etc..The temperature of the thermostat bath is maintained by liquid nitrogen and a heating coil.The temperature fluctuation in the bath is less than ±5 mK?(30 min)?1.The temperature is measured by a four-head 25-platinum resistance thermometer(Model: WZPB-I,China) with an uncertainty of ±1 mK(ITS) and a Keithley 2002 data acquisition/switch unit.The overall uncertainty of the measured temperature is within ±10 mK.The pressure is measured by a pressure transducer(Model: PMP4010,Drunk).The total uncertainty of the pressure measurement is within ±330 Pa.In addition,based on Visual BASIC,a measurement and control program is also developed to get on-time data logging.The experimental VLE data of the methane and nitrogen mixture obtained by using the developed apparatus were compared with the corresponding reference data,and they have a good agreement with each other.The operating results show that the experimental apparatus developed runs well.