The Mechanism of Errors Generated from Slug Inside Pinhole in Pressure Measurement

Weiting Liu,Jialu Li,Liang Hu,Rui Su,Xin Fu
DOI: https://doi.org/10.2139/ssrn.4365973
2023-01-01
Abstract:Pinholes are holes drilled perpendicularly to the surface, and manometric systems are connected behind. If the wall is hydrophilic, menisci form inside the pinhole. The path from the wall surface to the sensor die will contain both gas and liquid slug serving as a hybrid pressure-transmitting media. This article sheds light on the measurement errors resulting from the heterogeneous pressure-transmission media. The errors are explored by analyzing dynamic performance of menisci. It is found that surface tension and liquid viscosity act as spring and damper respectively, showing mass-spring-damper characteristics. There is also a phase lag when pressure wave passes through the two-phase pressure transmitting medium, and for a fixed pinhole the resonant frequency depends on the slug length. The result shows that the pinhole pressure readings deviate tremendously from the true pressure value at certain frequency band, and error arisen from slug is underestimated and little attentions are paid yet.
What problem does this paper attempt to address?