Uncertainty Analysis of Slot Die Coater Gap Width Measurement by Using A Shear Mode Micro-Probing System

So Ito,Yuan-Liu Chen,Yuki Shimizu,Hirotaka Kikuchi,Wei Gao,Kazuhiko Takahashi,Toshihiko Kanayama,Kunmei Arakawa,Atsushi Hayashi
DOI: https://doi.org/10.1016/j.precisioneng.2015.09.016
2016-01-01
Abstract:A shear mode micro-probing system was constructed for gap measurement of a precision slot die coater with a nominal gap width of 90 mu m and a length of 200 mm. A glass micro-stylus with a nominal tip ball diameter of 52.6 mu m was oscillated by a tuning fork quartz crystal resonator with its oscillation direction parallel to the measurement surfaces. An on-line qualification setup was established to compensate for the influences of the uncertainty sources, including the water layers on the measurement surfaces. The measurement uncertainty of the measured gap width was estimated to be less than 100 nm. (C) 2015 Elsevier Inc. All rights reserved.