Abstract:We studied and improved the performance of a terahertz (THz) direct detector based on the superconducting NbN hot electron bolometer (HEB) integrated with microwave readout for imaging arrays and other applications. A 6 nm thick NbN microbridge at the feed point of a twin-slot antenna is embedded into a high-Q quarter-wavelength coplanar waveguide (CPW) resonator made from a 350 nm thick NbN film which is weakly coupled to a CPW feedline. THz direct detection was performed at a signal frequency of 0.66 THz at a bath temperature of 4.2 K. When operating at a low input microwave power level, the device acting as a microwave kinetic inductance detector with a kinetic inductance fraction of 0.45 changes the surface impedance of the resonator in response to incident THz radiation. The estimated optical noise equivalent power (NEP) is similar to 0.23 pW center dot Hz(-0.5) which shows about one order of magnitude improvement compared to that of the traditional voltage-biased NbN HEBs. The improved performance of the THz direct detectors with lower NEP and easy readout makes them promising in imaging with a large number of the array pixels.