High Accuracy Measurement Instrument For Linear Thermal Expansion Coefficient

Yp Li,Zl Ding,F Yuan,Z Chen
2004-01-01
Abstract:A high accuracy measuring system for linear thermal expansion coefficient of large structure is described. The length measurement system employs a BP5529A dual frequency laser interferometer. The measurement process is based on the differential principle, which can avoid much adverse influence on the result of measurement such as the influences of force deformation and displacement of specimen. It fully explores the accuracy potentialities of a laser interferometer and guarantees a high enough measurement precision. A special measurement bracket system is developed, which can move along measurement axis, and suitable for measuring large structure. High precise elasticity support that without friction, gap, breakforce and backpath error is used to ensure high precise measurement. Platinum resistance thermometer that is excited by pulse electric current is used to measure temperature. With small volume, the platinum resistance can be easily sticked on specimen. Pulse electric current exciting lessens the heat caused by platinum resistances themselves, and improves the accuracy of temperature measurement. This measurement instrument can measure specimen up to 1m and the uncertainty is better than 5 x 10(-8)/degreesC.
What problem does this paper attempt to address?