Analysis of verification results of intensity-modulated radiation therapy for 2887 cases and its effect in quality control

刘思伟,赵波,李雅钢,汪春林,于世平,周刚
DOI: https://doi.org/10.3969/J.ISSN.1672-8270.2018.06.019
2018-01-01
Abstract:Objective: To analyze the reason that verification plan could not be approved and its effect in the quality control according to the verification result of intensity-modulated radiation therapy (IMRT). Methods:The intensity modulated plans of 2887 patients pre-radiotherapy were carried out verification analysis. These non-approved intensity modulated plans were carried out classified statistic as type of error, and the reasons were counter calculated according to statistical discrepancy and quality control results of accelerator. Results: 2838 cases of the 2887 cases were passed, and the qualified rate was 98.30%. Under the quality control of same standard, the qualified rates of one accelerator pre and post resurveying data were 96.92% and 99.24%, respectively. And the amount of non-approved plans, which reason was that the optimal result of single treatment planning system (TPS) was poor, were 16.33% in all of non-approved intensity-modulated plans. Conclusions: Based on ensuring the consistency of calculated dose of TPS and exposure dose of patients, the verification of IMRT pre-radiotherapy can more contribute to the comprehensive quality control for accelerator and radiotherapy process.
What problem does this paper attempt to address?