Experimental Verification of HW-Plan Treatment Planing System

CHEN Yong,BAO Shang-lian,HUANG Fei-zeng,YUAN Ke-hong,XIAO Gui-ping
DOI: https://doi.org/10.3969/j.issn.1005-202x.2005.06.002
2005-01-01
Abstract:The experiment proves it practicable that film dosimetry combined with ionization chamber measurement can be used to verify Treatment Planning System(TPS) in radiotherapy.The Recommendations of AAPM Radiation Therapy Committee Task Group 55 are adopted as the acceptance criteria for the comparison of point absolute doses,dose profiles along X and Y axis,and isodose curves between measured and calculated by the testing TPS.In experiment,dose distributions of target are measured with Kodak EDR2 radiographic films and the innovative customer-designed processing software while absolute point doses at isocenter and other critical points with ionization chamber under the experimental conditions that three slanting fields crossirradiate the target with the isoceter as their focus.Homogeneous solid phantom and water phantom,the image data and electronical densities of which are obtained by scanning with GE CT scanner in advance,are also used in this experiment.
What problem does this paper attempt to address?