Step and Shoot Planning of Intensity-Modulated Radiotherapy and 30 Patients Dosimetric Verification Analysis

Bai Penggang,Zhang Xiuchun,Pan Jianji,Li Huiling,Zhang Chun,Zhang Yu
DOI: https://doi.org/10.3969/j.issn.1672-4194.2006.03.014
2006-01-01
Abstract:Objective To discuss the process of step and shoot intensity-modulated radiotherapy, dosimetric verification and quality control. MethodsThirty patients were treated with step and shoot intensity-modulated radiotherapy(IMRT) technique from May 2003 to May 2004, including nasopharynx cancer, rectum cancer, lung cancer and so on. The plans were made in Nuclear PLATO planning treatment system. The doses of the interesting points and isodose distributions of the interesting planes in phantom were verified in Simens Primus accelerator. Results The error between the measured and calculated in the interesting points was less than 5%. Isodose distribution and fluent map by the film were consistent with those predicted by the planning. Conclusion The dosimetric verification of IMRT is a reliable measure in the course of radiotherapy.
What problem does this paper attempt to address?