Can Standard Radiation Therapy Quality Assurance (qa) Detect Potential Delivery Errors?
L. Ren,J. O'Daniel,J. Adamson,H. Yan,F. Yin
DOI: https://doi.org/10.1016/j.ijrobp.2012.07.2084
2012-01-01
Abstract:Purpose/Objective(s)To determine whether introduced delivery errors in IMRT and VMAT treatments can be detected by a 3D diode array using conventional analysis techniques.Material/MethodsA 3D QA device with 2 orthogonal diode array planes was used for IMRT and VMAT QA. The diode spacing is 5mm in the central 6cmx6cm region and 1cm in the peripheral 20cmx20cm region. 3D volumetric dose is interpolated from the measured planar dose. Patient plans were modified using an in-house program to introduce the following three types of errors in IMRT and VMAT delivery: (1). constant positioning errors for all leaves on one leaf bank (1, 2, and 3mm); (2). constant positioning errors for one central leaf (1, 3, and 20mm); (3). errors in collimator angle (2 and 5 deg). The measurements for the modified plans were compared to the calculated dose for the original plan for 3D gamma analysis to evaluate the sensitivity of the device to these delivery errors. Brain, head-and-neck, and prostate patient plans were tested for each type of error.ResultsThe Table shows the gamma analysis pass rates using different criteria (3%, 3mm or 2%, 2mm) for the original plan and the plan with introduced errors in IMRT and VMAT delivery. As shown in the table, when using the traditional 3%, 3mm criteria for gamma analysis, a 2mm positioning error of the leaf bank leads to obvious drop of the pass rate in both IMRT and VMAT QAs. However, a single leaf positioning error of up to 3mm and collimator rotation angle errors of up to 5 degrees didn't significantly reduce the pass rate. Large single leaf positioning errors of 20mm reduced the pass rate by a similar amount as the 2mm leaf bank positioning error. Reducing the gamma analysis criteria to 2mm, 2% enhanced the sensitivity of the device to delivery errors, especially errors in collimator rotation angles. Results of the 3D gamma map indicated that the failure region was mainly in the center of the field for leaf bank positioning errors, in the single leaf motion region for single leaf positioning errors, and in the boundary of the field for collimator rotation angle errors.ConclusionPoster Viewing Abstract 3491; Table3D gamma analysis pass rate for different delivery errors in IMRT and VMATIMRT, gamma analysis (3%, 3mm)IMRT, gamma analysis (2%, 2mm)VMAT, gamma analysis (3%, 3mm)VMAT, gamma analysis (2%, 2mm)Original plan100.0%100.0%97.8%97.2%Leaf bank 1mm99.7%95.2%99.2%94.5%Leaf bank 2mm95.9%78.9%90.5%76.5%Leaf bank 3mm77.4%56.7%75.0%57.0%Central leaf 1mm100.0%99.8%100.0%98.1%Central leaf 3mm100.0%99.1%99.7%96.1%Central leaf 20mm94.1%92.1%90.9%85.0%Collimator 2degree100.0%99.0%100.0%98.0%Collimator 5degree99.3%91.9%100.0%97.1% Open table in a new tab Purpose/Objective(s)To determine whether introduced delivery errors in IMRT and VMAT treatments can be detected by a 3D diode array using conventional analysis techniques. To determine whether introduced delivery errors in IMRT and VMAT treatments can be detected by a 3D diode array using conventional analysis techniques. Material/MethodsA 3D QA device with 2 orthogonal diode array planes was used for IMRT and VMAT QA. The diode spacing is 5mm in the central 6cmx6cm region and 1cm in the peripheral 20cmx20cm region. 3D volumetric dose is interpolated from the measured planar dose. Patient plans were modified using an in-house program to introduce the following three types of errors in IMRT and VMAT delivery: (1). constant positioning errors for all leaves on one leaf bank (1, 2, and 3mm); (2). constant positioning errors for one central leaf (1, 3, and 20mm); (3). errors in collimator angle (2 and 5 deg). The measurements for the modified plans were compared to the calculated dose for the original plan for 3D gamma analysis to evaluate the sensitivity of the device to these delivery errors. Brain, head-and-neck, and prostate patient plans were tested for each type of error. A 3D QA device with 2 orthogonal diode array planes was used for IMRT and VMAT QA. The diode spacing is 5mm in the central 6cmx6cm region and 1cm in the peripheral 20cmx20cm region. 3D volumetric dose is interpolated from the measured planar dose. Patient plans were modified using an in-house program to introduce the following three types of errors in IMRT and VMAT delivery: (1). constant positioning errors for all leaves on one leaf bank (1, 2, and 3mm); (2). constant positioning errors for one central leaf (1, 3, and 20mm); (3). errors in collimator angle (2 and 5 deg). The measurements for the modified plans were compared to the calculated dose for the original plan for 3D gamma analysis to evaluate the sensitivity of the device to these delivery errors. Brain, head-and-neck, and prostate patient plans were tested for each type of error. ResultsThe Table shows the gamma analysis pass rates using different criteria (3%, 3mm or 2%, 2mm) for the original plan and the plan with introduced errors in IMRT and VMAT delivery. As shown in the table, when using the traditional 3%, 3mm criteria for gamma analysis, a 2mm positioning error of the leaf bank leads to obvious drop of the pass rate in both IMRT and VMAT QAs. However, a single leaf positioning error of up to 3mm and collimator rotation angle errors of up to 5 degrees didn't significantly reduce the pass rate. Large single leaf positioning errors of 20mm reduced the pass rate by a similar amount as the 2mm leaf bank positioning error. Reducing the gamma analysis criteria to 2mm, 2% enhanced the sensitivity of the device to delivery errors, especially errors in collimator rotation angles. Results of the 3D gamma map indicated that the failure region was mainly in the center of the field for leaf bank positioning errors, in the single leaf motion region for single leaf positioning errors, and in the boundary of the field for collimator rotation angle errors. The Table shows the gamma analysis pass rates using different criteria (3%, 3mm or 2%, 2mm) for the original plan and the plan with introduced errors in IMRT and VMAT delivery. As shown in the table, when using the traditional 3%, 3mm criteria for gamma analysis, a 2mm positioning error of the leaf bank leads to obvious drop of the pass rate in both IMRT and VMAT QAs. However, a single leaf positioning error of up to 3mm and collimator rotation angle errors of up to 5 degrees didn't significantly reduce the pass rate. Large single leaf positioning errors of 20mm reduced the pass rate by a similar amount as the 2mm leaf bank positioning error. Reducing the gamma analysis criteria to 2mm, 2% enhanced the sensitivity of the device to delivery errors, especially errors in collimator rotation angles. Results of the 3D gamma map indicated that the failure region was mainly in the center of the field for leaf bank positioning errors, in the single leaf motion region for single leaf positioning errors, and in the boundary of the field for collimator rotation angle errors. ConclusionPoster Viewing Abstract 3491; Table3D gamma analysis pass rate for different delivery errors in IMRT and VMATIMRT, gamma analysis (3%, 3mm)IMRT, gamma analysis (2%, 2mm)VMAT, gamma analysis (3%, 3mm)VMAT, gamma analysis (2%, 2mm)Original plan100.0%100.0%97.8%97.2%Leaf bank 1mm99.7%95.2%99.2%94.5%Leaf bank 2mm95.9%78.9%90.5%76.5%Leaf bank 3mm77.4%56.7%75.0%57.0%Central leaf 1mm100.0%99.8%100.0%98.1%Central leaf 3mm100.0%99.1%99.7%96.1%Central leaf 20mm94.1%92.1%90.9%85.0%Collimator 2degree100.0%99.0%100.0%98.0%Collimator 5degree99.3%91.9%100.0%97.1% Open table in a new tab