Quantification of Treatment Plan Deliverability in Breast Volumetric-modulated Arc Therapy With Agility Multi-leaf Collimator

DO Hoon Oh,Jin Hwa Choi,Hyejo Ryu,Minsoo Chun
DOI: https://doi.org/10.21873/invivo.13690
In Vivo
Abstract:Background/aim: The aim was to assess the complexity of breast volumetric-modulated arc therapy (VMAT) plans using various indices and to evaluate their performance through gamma analysis in predicting plan deliverability. Materials and methods: A total of 285 VMAT plans for 260 patients were created using the VersaHD™ linear accelerator with a Monaco treatment planning system. Corresponding verification plans were generated using the ArcCHECK® detector, and gamma analysis was conducted employing various criteria. Twenty-eight plan complexity metrics were computed, and Pearson's correlation coefficients were determined between the gamma passing rate (GPR) and these metrics. Results: The average GPR values for all plans were 97.7%, 89.9%, and 78.0% for the 2 mm/2%, 1 mm/2%, and 1 mm/1% criteria, respectively. While most complexity metrics exhibited weak correlations with GPRs under the 2 mm/2% criterion, leaf sequence variability (LSV), plan-averaged beam area (PA), converted area metric (CAM), and edge area metric (EAM) demonstrated the most robust performance, with Pearson's correlation coefficients of 0.57, 0.50, -0.70, and -0.56, respectively. Conclusion: Metrics related to beam aperture size and irregularity, such as LSV, PA, CAM and EAM, proved to be reasonable predictors of plan deliverability in breast VMAT.
What problem does this paper attempt to address?