From plan to delivery: Machine learning based positional accuracy prediction of multi‐leaf collimator and estimation of delivery effect in volumetric modulated arc therapy

Minmin Qiu,Jiajian Zhong,Zhenhua Xiao,Yongjin Deng
DOI: https://doi.org/10.1002/acm2.14437
2024-06-22
Journal of Applied Clinical Medical Physics
Abstract:Purpose The positional accuracy of MLC is an important element in establishing the exact dosimetry in VMAT. We comprehensively analyzed factors that may affect MLC positional accuracy in VMAT, and constructed a model to predict MLC positional deviation and estimate planning delivery quality according to the VMAT plans before delivery. Methods A total of 744 "dynalog" files for 23 VMAT plans were extracted randomly from treatment database. Multi‐correlation was used to analyzed the potential influences on MLC positional accuracy, including the spatial characteristics and temporal variability of VMAT fluence, and the mechanical wear parameters of MLC. We developed a model to forecast the accuracy of MLC moving position utilizing the random forest (RF) ensemble learning method. Spearman correlation was used to further investigate the associations between MLC positional deviation and dosage deviations as well as gamma passing rates. Results The MLC positional deviation and effective impact factors show a strong multi‐correlation (R = 0.701, p‐value 0.05). Conclusions The RF predictive model provides a prior tool for VMAT quality assurance.
radiology, nuclear medicine & medical imaging