Impact of accelerator′s multi-leaf collimator leaves on the intensity modulated radiation therapy dose at different gantry angles
Chang-li RUAN,Li-ming XU,Qi-bin SONG,Guang-dong MA,Chang-hu LI
DOI: https://doi.org/10.3760/cma.j.issn.1004-4221.2011.04.024
2011-01-01
Abstract:Objective To investigate the impact of accelerator′s multi-leaf collimator (MLC) on the radiotherapy dose with different gantry angles.Methods Measured with Mapcheck 2D diode array and 30 cm×30 cm×3 cm solid water, Pre-selecting the 30 appropriate single fields and 0°,45°,90°,270°,315° gantry angles of static and dynamic intensity modulated radiation therapy (IMRT), quantification analysis of the passing rate with MapCheck γ(3%/3 mm) and (5%/3 mm) analysis methods, and the same method to 30 examples static and dynamic IMRT plans.Results When the accelerator collimator angle is 0°,the 30 appropriate single fields′ passing rate of between 0°gantry angle and 45°,90°,270°,315°gantry angles of static and dynamic IMRT accordingly is 97.71% and 96.25%(t=1.70, P=0.389), 96.34% and 93.72%(t=2.95, P=0.002), 96.65% and 92.98%(t=2.87, P=0.005), 95.87% and 93.15%(t=2.71, P=0.006), 96.09% and 93.51%(t=2.89, P=0.004) with MapCheck γ(3%/3 mm) analysis methods, however, the passing rate also does not have the difference, respectively is 99.31%-99.73% and 98.89-99.68%(t=0.57-1.90, P=0.913-0.725) with MapCheck γ(5%/3 mm) analysis methods;the passing rate of 30 examples static and dynamic IMRT plans accordingly is 96.11%-96.76% and 94.88%-95.78%(t=1.02-1.61, P=0.317-0.235).Conclusions When the accelerator collimator angle is 0°, at different gantry angles, MLC leaves due to gravity, friction, inertia, etc caused by errors in place, the physical penumbra and leakage radiation will indeed affect the IMRT dose, however, the deviation of dose distribution is within acceptable 5%.