Effect of the Collimator Angle on Dosimetric Verification of the Volumetric Modulated Arc Therapy

Yong Ho Kim,Ha Ryung Park,Won Taek Kim,Dong Won Kim,Yongkan Ki,Juhye Lee,Jinsuk Bae,Dahl Park
DOI: https://doi.org/10.48550/arXiv.1503.03569
2015-03-12
Abstract:Collimator angle is usually rotated when planning volumetric modulated arc therapy (VMAT) due to the leakage of radiation between multi-leaf collimator (MLC) leaves. We studied the effect of the collimator angles on the results of dosimetric verification of the VMAT plans for head and neck patients. We studied VMAT plans for 10 head and neck patients. We made 2 sets of VMAT plans for each patient. Each set was composed of 10 plans with collimator angles of 0, 5, 10, 15, 20, 25, 30, 35, 40, 45 degrees. Plans in the first set were optimized individually and plans in the second set shared the 30 degree collimator angle optimization. Two sets of plans were verified using the 2-dimensional ion chamber array MatriXX (IBA Dosimetry, Germany). The comparison between the calculation and measurements were made by the $\gamma$-index analysis. The $\gamma$-index (2\%/2 mm) and (3\%/3 mm) passing rates had negative correlations with the collimator angle. Maximum difference between $\gamma$-index (3\%/3 mm) passing rates of different collimator angles for each patient ranged from 1.46\% to 5.60\% with an average of 3.67\%. There were significant differences (maximum 5.6\%) in the passing rates of different collimator angles. The results suggested that the accuracy of the delivered dose depends on the collimator angle. These findings are informative when choosing a collimator angle in VMAT plans.
Medical Physics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is: to study the influence of different collimator angles on the dose verification results of volumetric modulated arc therapy (VMAT) plans. Specifically, by analyzing the dose verification data at different collimator angles, the author explored how the collimator angle affects the dose delivery accuracy of VMAT plans, especially for the treatment plans of head - and - neck cancer patients. In VMAT treatment, the selection of the collimator angle is an important parameter because it can reduce the radiation leakage between the multi - leaf collimator (MLC) leaves, thereby improving the control of the dose distribution. However, different collimator angles may lead to differences in the accuracy of dose delivery. Therefore, understanding and quantifying these differences are crucial for optimizing VMAT plans. The paper experimentally studied the VMAT plans of 10 head - and - neck cancer patients at different collimator angles, used the two - dimensional ionization chamber array MatriXX for dose verification, and compared the differences between the calculated values and the measured values through gamma - index analysis. The study found that as the collimator angle increases, the pass rates of the gamma - index (2%/2 mm) and (3%/3 mm) are negatively correlated, that is, the larger the collimator angle, the lower the pass rate of dose verification. This indicates that selecting an appropriate collimator angle is very important for improving the dose delivery accuracy of VMAT plans. According to the research results, the author suggests that in the head - and - neck VMAT plans in their hospital, the collimator angle should be selected between 15° and 25° to ensure a higher dose verification pass rate.