Commissioning of a commercial treatment planning system for scanned carbon‐ion radiotherapy

Wei Sun,Weiwei Wang,Zhijie Huang,Jingfang Zhao
DOI: https://doi.org/10.1002/acm2.14580
2024-11-30
Journal of Applied Clinical Medical Physics
Abstract:Purpose To commission the RayStation (RS) TPS (treatment planning system) for scanned CIRT (carbon‐ion radiotherapy) utilizing pencil beam algorithms (PBv4.2). Methods The beam model commissioning entailed employing 1D single beams and 2D monoenergetic fields to validate spot profiles with films, assess beam range using Peakfinder measurements, and evaluate fragment spectra through dose‐averaged linear energy transfer (LETd) calculations. 3D dose distributions were verified in homogeneous phantoms for both absorbed and relative biological effectiveness (RBE)‐weighted doses, and further assessed in double wedge and anthropomorphic phantoms for absorbed dose only. Finally, RBE‐weighted dose verification and patient‐specific quality assurance were conducted using 58 beams from 20 clinically treated patient plans. Results The results demonstrated good agreement in absolute dose distribution between TPS calculations and measurements, with mean dose discrepancies within 3%. However, deviations were slightly higher (> 1%) for the cases involving the range shifter (RaShi) compared to those without the RaShi (
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?