Treatment planning system commissioning of the first clinical biology‐guided radiotherapy machine

Eric Simiele,Dante Capaldi,Dylan Breitkreutz,Bin Han,Timothy Yeung,John White,Daniel Zaks,Michael Owens,Srinath Maganti,Lei Xing,Murat Surucu,Nataliya Kovalchuk
DOI: https://doi.org/10.1002/acm2.13638
2022-05-31
Journal of applied clinical medical physics [electronic resource] / American College of Medical Physics
Abstract:The RefleXion X1 is a novel radiotherapy machine designed for image-guided radiotherapy (IGRT) and biology-guided radiotherapy (BgRT). Its treatment planning system (TPS) generates IMRT and SBRT plans for a 6MV-FFF beam delivered axially via 50 firing positions with the couch advancing every 2.1 mm. The purpose of this work is to report the TPS commissioning results for the first clinical installation of RefleXionTM X1.