Quality assurance and quality control of IGRT system

胡彩容,张秀春,吴君心,潘建基
DOI: https://doi.org/10.3969/j.issn.1672-8270.2010.02.003
2010-01-01
Abstract:Objective:With the accumulated experience from the quality assurance and quality control work that our institute has done since the beginning of IGRT technology,the major quality control and monitoring programs and realization methods in IGRT system are analyzed and discussed,to ensure the safe,efficient and stable operation.Methods:The image low contrast visibility, spatial resolution,uniformity and the geometry accuracy of the imaging system were detected,the registration accuracy of the system was tested,the isocenter of the six-degree couch was calibrated.Results:The image low contrast visibility was below 2%,the spatial resolution was no less than 7lp/ cm,the deviation of the pixel value which indicates the image uniformity was below 2%,the geometry accuracy which represents the degree of image distortion was below 1.04mm in X,Y,Z directions, respectively.When the XVI system rotates with the gantry,the deviation of rotation isocenter due to gravity was below 4pixel;the deviation between the KV system and MV system was below 0.5mm in X, Y,Z directions,respectively.The deviation between the six-degree couch and the isocenter was within 0.2mm,0.2°.Conclusion:The corresponding quality control programs and methods could ensure the operational safety and stability,and helped to minimizing the setup errors during the patients' radiotherapy.