Analysis of Occluder Displace after Transcatherter Occlusion of Ventricular Septal Defect

ZHANG Yu-shun,LI Huan,DAI Zheng-xue,LI Jun,ZHANG Jun,WANG Xiao-yan
DOI: https://doi.org/10.3969/j.issn.1009-7236.2005.02.026
2005-01-01
Abstract:AIM: To investingate the reasons of occluder displace and its preventive measures after transcatheter occlusion of ventricular septal defect (VSD). METHODS: 4 (3 male, 1 female) cases happened occluder displace after transcatheter occlusion of VSD. Of whom 3 cases were peri-membranous VSD with membranous ventricular septal aneurysm, 1 case was VSD. The diameter of defects were 13, 12, 12, and 10 mm, respectively. The implanting occluders were 16 mm, 12 mm domestic-made concentric VSD occluder, 14 mm Amplatzer VSD occluder, and 14 mm domestic-made eccentric VSD occluder, respectively. RESULTS: The four cases were all successfully implanted with once try, and had no residual shunt immediately after occlusion. 2 cases happened haemolysis in the 4 h and 12 h of post-occlusion, and both occluders position were not found significant change, but both had significant residual shunt. After treated with drugs, both haemolysis were recovery and residual shunt were disappeared. 1 case in the 4 d of post-occlusion, ECG found third-degree atrioventricular block and echocardiography found occluder obviously displaced to the outlet of membranous aneurysm. The case had to be transferred to surgically remove occluder and repair VSD, and third-degree atrioventricular block was recovered after surgery. 1 case with VSD, 2 weeks after occlusion, echocardiography found occluder was displaced to the right ventricular and had obviously residual shunt. 4 months follow-up, residual shunt examined by echocardiography was diminished. CONCLUSION: Transcatheter occlusion of VSD, occluder displace is associated with VSD pathological anatomy. Occluder displace mostly happened in the peri-membranous VSD with membranous ventricular septal aneurysm and VSD.
What problem does this paper attempt to address?