Transcatheter Closure of Secundum Atrial Septal Defect with Amplatzer Occluder Device

王胜煌,陈君柱,张芙荣
DOI: https://doi.org/10.3969/j.issn.1001-5132.2002.02.034
2003-01-01
Abstract:Objective To evaluate the efficiency of transcatheter closure of secundum atrial septal defect (ASD) with Amplatzer occluder device.Methods Twenty-one patients with ASD was treated using Amplatzer occluder device through the percutaneous procedure under fluoroscopy and transesophageal echocardiography (TEE) guidance.Results The mean diameter of ASD was 20.5 ± 7.1mm (ranged from 6 to 32mm)shown by TEE before the procedure,The mean balloon stretched diameter of the ASD was 23.3±7.7mm (ranging from 9 to 34 mm).The mean diameter of the Amplatzer occluder selected was 25.9±8.1mm (ranging from 10 to 38 mm).For two cases with double ASD,Two occluder device of 22 and 10 mm in diameter were used in one patient,a larger occluder device of 28 mm in diameter was used for closing two defects in the other patient with good result.All patients were treat successfully.The success rate of placement of the Amplatzer occluder was 100%.No complications occurred.All of 21 patients was no residual shunt found during 1~18 months follow-up.Conclusion Transcatheter closure of secundum ASD using Amplatzer septal occluder device was an efficient and safe nonsurgical method.
What problem does this paper attempt to address?