Echocardiographic Guidance of Transcatheter Closure of Atrial Septal Defect

温朝阳,李越,王廉一,王广义,陈练,王禹
DOI: https://doi.org/10.3760/j.issn:1004-4477.1999.05.008
1999-01-01
Abstract:Objective This study reports our experience of echocardiographic guidance with transcatheter closure of atrial septal defects(ASD)in patients,using the Amplatzer,a new occlusion device.Methods Nineteen patients with secundum ASD waiting for surgical treatment met transesophageal echocardiography(TEE)[3 transthoracic echocardiography(TTE)]for transcatheter closure.The Amplatzer size was chosen to be equal to or 1~2mm more than the stretched diameter of ASD.Under uhrasound guidance,The device was advanced to ASD with guiding sheath,and then its left disk,waist and right disk was deployed in turn,Once its position was optimal,it was released.Results Sixteen patients were treated successfully except 3 patients with:>34mm stretched diameters,which were too large for the Amplatzers.The mean ASD diameter by echcardiography was significantly smaller than the mean stretched diameter of the ASD[(18.06±7.00)mm vs(22.56±6.25)mm].The devices'positions were optimal and no residual shunt was found by TTE immediately,24 h,1 month and 3 months after the procedure.NO complication was observed during the 3 months fellow up. Conclusions Echocardiography was very important in the selectign of patient,the guidance of the device deployed and the follow-up in the transcatheter closure of ASD with the Amplatzer.