Calibration for the 3D dental cast scanner system

Qing YU,Yong WANG,Peijun LV
DOI: https://doi.org/10.3969/j.issn.1003-7632.2005.02.012
2005-01-01
Abstract:Objective To calibrate the 3D dental cast scanner system and to evaluate the reliability of the calibration method.Methods Two experiments were included in this study. ① A fixed-point scaling experiment with a custom scaling block. ②A comparison of the measured data from dental casts with a mechanical 3D digitizing system (YM-2115) vs. manual measurement.Results After calibration, there were no significant differences among the dental cast data acquired with the D.02-L-3D SCANNER?YM-115 and the manual method.Conclusion The calibration method is feasible and reliable for routine adjustments of the D.02-L-3D SCANNER.