Effect of a calibration aid and the intraoral scanner on the registration of a partially edentulous maxilla: An in vitro study

Sophia Boehm,Stefan Rues,Alexandra Balzer,Peter Rammelsberg,Moritz Waldecker
DOI: https://doi.org/10.1016/j.prosdent.2024.02.019
IF: 4.148
2024-03-14
Journal of Prosthetic Dentistry
Abstract:Statement of problem Precise data are a prerequisite for accurately fitting restorations. Therefore, intraoral scanners have to be reliable. Data on differences between the same model of intraoral scanning systems are lacking. Purpose This in vitro study evaluated differences in the scanning accuracy of a partially edentulous maxilla between combinations of new intraoral scanners of a single model from the same manufacturer (Primescan; Dentsply Sirona) and several calibration aids. Material and methods The in vitro reference model simulated a maxilla with 6 prepared teeth for a complete arch fixed partial denture. Five precision ceramic balls were used to detect dimensional deviation between the reference model and the scan. Distances were divided into 4 categories, from short distances between 2 neighboring precision balls to the cross-arch distance with the scan path comprising all 5 balls. For each combination of 4 new intraoral scanners and their respective calibration aids, 12 model registrations were generated. The data were statistically analyzed using ANOVA (α=.05). Results Distance deviations increased with increasing scan path length and were significantly affected by the covariates "scanner" ( P ≤.023) and, for 3 of 4 distance categories, the "calibration aid" (short, medium, and long distances: P ≤.013). For short and medium distances, acceptable scanning results were achieved for all test groups. The largest deviation was 539 μm for the cross-arch distance. Conclusions Scanning accuracy depends on the scanner and the calibration aid used, in particular, for spans exceeding a single quadrant.
dentistry, oral surgery & medicine
What problem does this paper attempt to address?