Comparison study of the imaging quality ofSE-EPI-DWI and STIR-DWI in breast lesions

林婷婷,董江宁,邓克学,韦炜,方昕,韦超,韦树华,张萍
DOI: https://doi.org/10.19405/j.cnki.issn1000-1492.2017.08.019
2017-01-01
Abstract:Objective To compare the image quality and ADC values of SE-EPI-DWI and STIR-DWI in breast lesions.Methods Magnetic resonance imaging(MRI) images of 40 patients(both SE-EPI-DWI and STIR-DWI performed) were retrospectively analysed.Parameters compared between SE-EPI-DWI and STIR-DWI were image artefacts, image signal-to-noise(SNR),contrast-to-noise(CNR),and apparent diffusion coefficient(ADC).Results The rates of image artefacts was significantly higher for SE-EPI-DWI(P<0.001):quality levels 0, 1, 2, and 3(best) accounted for 0, 7.5%, 27.5% and 65.0% of STIR-DWI iamges, and 12.5%, 25.0%, 30.0% and 32.5% of SE-EPI-DWI iamges,respectively.Excepted for 5 cases of unqualified images,the SNR and CNR were significantly lower of STIR-DWI images(P<0.001).The ADC values of breast leisons were not significantly different.Conclusion The SNR and CNR of SE-EPI-DWI is higher than STIR-DWI, while the image artefacts of STIR-DWI is less than SE-EPI-DWI.STIR-DWI is an important supplementary sequence for clinical use in breast MRI investigations.Combine the two sequences can improve the success rate for breast DWI.