Study of Glass Hydrometer Calibration by Hydrostatic Weighting

Chao-yun CHEN,Jin-tao WANG,Zhi-hao LI,Bin SUN
DOI: https://doi.org/10.3969/j.issn.1000-1158.2017.02.16
2017-01-01
Abstract:Glass hydrometers calibration uses silicon ring as a reference standard solid density, tri-tretradecane with density stability and low surface tension as the standard working liquid, CCD image measurement system to align the scale of hydrometer and liquid surface, designes a glass hydrometer calibration system.According to twice glass hydrometer weighing in the air and liquid can calculate the correction value of the current scale.To measure the density range of 770~790 kg/m3, a resolution of 0.2 kg/m3 of hydrometer is used.The results of measurement comparing with that of the Physikalisch-Technische Bundesanstalt,verify the validity of the calibration system.
What problem does this paper attempt to address?