Measurement Results Comparison Between Laser Particle Analyzer and Sieving Method in Particle Size Distribution

LI Wen-kai,WU Yu-xin,HUANG Zhi-min,FAN Rong,LV Jun-fu
DOI: https://doi.org/10.3969/j.issn.1008-5548.2007.05.004
2007-01-01
Abstract:In order to compare the difference between the Malvern laser analyzer and the sieving method in particle size distribution measurement,as many as 16 samples of CFB fly ash were measured by both methods respectively.The results from different methods showed a great diversity.The percentages of ultrafine particles obtained with laser analyzer are overestimated.This is considered to be due to the shape of particles.On the other side,the sieving method showed great errors in the measurements of small-sized particles,but more precise in the measurement of bigger particles.Micrography of particles with different size region were taken,which proved that the conglomeration of small particles was the main reason why results were smaller in the measurements of small-sized particles with the sieving method.