Comparison Between Measurement Results from Two Types of Laser Particle Size Analyzers

肖尚斌,李安春,刘建国,徐兆凯
DOI: https://doi.org/10.3969/j.issn.1000-3096.2004.12.004
2004-01-01
Abstract:Grain size analysis for149samples was carried out with two types of laser size analyzers(Cilas940L and Malvern2000).Measurement data was analyzed and compared in different particle ranges and grain size parameters.This paper gives the correlation of the two types of laser particle analyzers in different size.The results show that the measurement between the two types of instruments is closest in13~14μm,and measurement result of Cilas940Lis bigger than that of Malvern2000in the range below13~14μm and smaller in the range above13~14μm.It is indicated that Cilas940L is more sensitive to finer particles than Malvern2000.