Determination of Pridinol Mesylate Injection by RP-HPLC

Haichun Zeng,Cao Jiang,Xiaohong Luo,Shuai Wang,Mingguo Tan,Fei Long,Yonghuang Luo
2014-01-01
Latin american journal of pharmacy
Abstract:The present study aims to develop a method for simultaneous determination of pridinol mesylate injection. The injection was analysed with reversed phase high performance liquid chromatography (RP-HPLC) under the following conditions: a Ultimate XB C18 (250 mm x 4.6 mm, 5 mu m) column, the mobile phase consisting of methanol and 0.1% phosphoric acid (7: 3, v/v) (containing 4 mmol/L sodium 1-octanesulfonate), the flow rate of 1.0 mL/min, the detection wavelength of 215 nm, the column temperature at 30 degrees C and the injection volume of 20 mu L. Under the chromatographic conditions, pridinol mesylate was completely separated from impurities. The calibration curve showed good linear correlation in the range of 5 similar to 300 mu g/mL (r(2) > 0.999). The average recovery was 97.83%. The minimal detection limit and the minimal quantification limit of pridinol mesylate were 0.17 and 3 ng, respectively. The method was simple, accurate, specific and is suitable for determination of pridinol mesylate injection.
What problem does this paper attempt to address?