Determination of Ziprasidone Hydrochloride by RP-HPLC with UV Detection
N Li,Shun-Qing Shen,Y Zhong,Zhenyu Wang,YQ Yu
DOI: https://doi.org/10.3969/j.issn.1672-8467.2004.06.030
2004-01-01
Abstract:Purpose: To establish an RP-HPLC method for the determination of Ziprasidone hydrochloride. Methods: A C18 (4.6 mm x 150 mm, 5 μm) column was adopted as stationary phase, Methanol-50 mmol/L sodium acetate solution (adjustment of pH to 6.0 with acetic acid) (75:25, V/V) was used as mobile phase, the flow rate was 1.2 mL/min, the detection wavelength was 254 nm, and the temperature of column was 25°C. Results: The limit of detection for Ziprasidone hydrochloride was 10.10 ng, the linearity was good at the range of 1.010-1010 μg/mL, R = 1.000. The intra-day and inter-day RSD of ziprasidone hydrochloride were both lower than 2%. Conclusions: The established method was simple, sensitive and accurate,and can be used for the determination and quality control of ziprasidone hydrochloride.