Ion-pair RP-HPLC simultaneous determination of four alkaloids in compound Kushen injection

CHEN Chen,QU Hai-bin,HAO Run-cai,DONG Mi-zhu,LI An-ping,SHAO Qing
DOI: https://doi.org/10.16155/j.0254-1793.2012.06.011
2012-01-01
Abstract:Objective:To establish an ion-pair RP-HPLC method for the simultaneous determination of matrine,sophocarpine,oxysophocarpine and oxymatrine contents in compound Kushen injection.Methods:The chromatographic separation was carried out with a Diamonsil C18 column(250 mm×4.6 mm,5.0 μm).The mobile phase was composed of solvent A(0.04%phosphoric acid and 10 mmol·L-1 sodium pentanesulfonate in water) and solvent B(acetonitrile) with linear gradient elution at a flow rate of 1.2 mL·min-1.The column temperature was 30 ℃.The detection wavelength was 210 nm.Results:The calibration curves had good linear for matrine,sophocarpine,oxysophocarpine,and oxymatrine in the range of 18.92-946.0 mg·L-1,4.052-202.6 mg·L-1,9.293-464.6 mg·L-1 and 34.73-1736 mg·L-1,respectively.The values of correlation coefficient were 0.9999.Repeatability experiments showed that RSD values of the 4 contents were all less than 2.1%.The average recoveries of matrine,sophocarpine,oxysophocarpine and oxymatrine were 98.2%,96.4%,96.9%,97.1%,respectively.Conclusion:The method is simple,accurate and good reproducibility,which can be used for quality control of Compound Kushen injection.
What problem does this paper attempt to address?