Effect of Very High-Fluence Proton Radiation on 6H-Sic Photoconductive Proton Detectors

Qing Liu,Dong Zhou,Xiaolong Cai,Ming Qi,Weizong Xu,Dunjun Chen,Fangfang Ren,Rong Zhang,Youdou Zheng,Hai Lu
DOI: https://doi.org/10.1109/led.2019.2949174
IF: 4.8157
2019-01-01
IEEE Electron Device Letters
Abstract:In this work, the effect of very high-fluence 100 MeV proton radiation on the performance of 6H-SiC photoconductive proton detectors is studied. The irradiation fluence is up to 1.6 x 10(17) cm(-2) and the degradation process of the SiC detector is continuously monitored for the first time. Before proton irradiation, the detector shows a low dark current of similar to 0.8 nA/cm(2) at 1000 V bias. As the irradiation fluence increases, the dark current continuously drops, which should be caused by irradiation damage induced carrier compensation defects. Meanwhile, the output current of the SiC detector shows an exponential decay behavior and tends to saturate at irradiation fluence up to 5 x 10(16) cm(-2). At the end of the very high fluence irradiation test, the detector still exhibits similar to 20% of its original output current value, suggesting the excellent radiation hardness of SiC for proton detection.
What problem does this paper attempt to address?