Experiment Report on Uric Acid Tracing by Sinew Bio-technology Test

CHEN Wei,ZHOU Shuai,HAN Shuai,DENG Zhi-wu,TAN Tao
DOI: https://doi.org/10.3969/j.issn.1672-9455.2011.21.013
2011-01-01
Abstract:Objective To create the traceability of Sinew Bio-technology test′s uric acid measurement system,and to improve the accuracy of customers′ final test results,and to contribute to the mutual recognition of the test results.Methods Sinew Bio-technology test′s own traceability flow chart was created on the basis of traceability chain map in the GB/T 21415-2008/ISO 17511:2003 of international standards;reference materials were purchased(NIST SRM 909b).Firstly,the manufacturers′ working calibrator was traced to reference material.Then the product calibrator was traced to uric acid working calibrator and the uncertainty of the combination was calculated,and to complete the tracing of the value of uric acid product calibrator made by Sinew Bio-technology test.Results Clinical comparisons were made by testing and fresh clinical serum samples,in order to ensure the interoperability of the reference materials.Meanwhile,SPSS17.0 was used to carry out statistical analysis the 1/4 CLIA′88 total allowable error and 97% prediction interval were set as the standard,in order to meet the requirement of value transferring verification,to confirm the uncertainty degree and complete the tracing job.Conclusion Sinew Bio-technology test successfully assigns value to the product calibrator due to the creation of self-built traceability process,achieves the tracing of product calibrator and improves the accuracy of the test results of reagents manufactured by Sinew Bio-technology test.