Establishment of Reference Method for Serum Urea Detection and Comparison with Routine Urea Kits

ZHANG Ya-jun,ZHANG Jie
DOI: https://doi.org/10.3969/j.issn.1006-7795.2011.02.022
2011-01-01
Abstract:Objective To establish and validate the coupled-enzyme equilibrium reference method as a reference method for urea detection and compare it with other 4 routine urea kits. Methods The reference method was verified by testing the two levels of urea in SRM909b and participating in RELA-2009.The EP9-A2 protocol was used to compare the results in forty serum samples in 4 routine kits methods with coupled-enzyme equilibrium reference method. Results The deviation of SRM909b level 1 was 2.19% and SRM 909b level 2 were 0.24% respectively through determining by the reference method.The results for RELA-2009 were within the accepted range.Comparing the confidence interval of the predicted bias with the definition acceptable error at the medical decision point Xc(Xc=9.64 mmol/L) for the 4 routine kits,the results were all within the range of defined acceptable error(4.5%) and the range of biological variation bias(5.5%). Conclusion The coupled-enzyme equilibrium reference method for serum urea was established in our lab.The urea results measured with 4 routine kits were different from results with the reference method,but the deviation was acceptable,and it will not affect the detection results.