Study on Traceability Scheme of Urea Measurement System of XC Test

CHEN Wei,ZHOU Shuai,HAN Shuai,DENG Zhi-wu,TAN Tao
DOI: https://doi.org/10.3969/j.issn.1672-9455.2011.17.026
2011-01-01
Abstract:Objective To establish the traceability of XC Test ′s urea measurement system to improve the accuracy of customers′ final test results and contribute to the mutual recognition of the test results.Methods To establish XC Test′s own traceability flow chart on the basis of traceability chain map in the international standards GB/T 21415-2008/ISO 17511:2003.To purchase the reference material(NIST SRM 909b).First,to trace manufacturers working calibrator to reference material.Then,to trace the product calibrator to urea working calibrator and to calculate the uncertainty of the combination,to complete the tracing of the value of urea product calibrator made by XC Test.Results By testing and comparing the fresh clinical serum samples,the interoperability of reference materials was ensured.Meanwhile,SPSS17.0 was adopted to conduct the statistical analysis.Under the standard of the 1/4CL IA′88 total allowable error of 97% prediction interval and testing projects,the requirement of value transferring verification was reached.The uncertainty was further determined till completing the tracing job.Conclusion XC Test succeeds in the value assignment of the product calibrator since the creation of self-built traceability process,which realizes the tracing of product calibrator and improves the accuracy of the test results of reagents manufactured by XC Test.
What problem does this paper attempt to address?