Traceability research on alanine aminotransferase measurement system

LUO Yu-xuan,FENG Chun-yan,WEI Jian-wei,CHEN Wei,HAN Shuai
DOI: https://doi.org/10.3969/j.issn.1672-9455.2012.11.005
2012-01-01
Abstract:Objective To create the traceability of alanine aminotransferase measurement system in the laboratory of our hospital for improving the accuracy of final testing results.Methods To create the self-build hospital test traceability flow chart on the basis of traceability chain map in the GB/T 21415-2008/ISO 17511:2003 of international standards;to purchase the reference material(IRMM ERM AD 454).First,the working calibrator provided by manufacturers was traced to reference material.Then the product calibrator was traced to manufactures working calibrator.The combination uncertainty was calculated for completing the tracing of the amount value of alanine aminotransferase product calibrator.The value assignment was135±5U/L.Results By testing and clinically comparing the fresh clinical serum samples,the interoperability of reference materials was ensured.Meanwhile,SPSS17.0 was adopted for statistical analysis.The total allowable error of the testing item 1/4CL IA′88 and 97% prediction interval were used as the standard,the requirement of amount value transferring verification was reached.The uncertainty was further determined to complete the tracing job.The valae assignment was expressed as(135±5)U/L.Conclusion The hospital traceability flow chart is successfully established.The product calibrator is performed the valuation correction.The accuracy of the test results of reagents is improved.