Optimal Choice of Collimation and Reconstructive Method in Spiral CT Portography

吴东,周康荣,陈祖望,陈刚,陈锦,陈惠明
DOI: https://doi.org/10.3969/j.issn.1001-9324.2000.01.012
2000-01-01
Abstract:Objective To optimize the collimation and the reconstructive method in spiral CT portography (CTP).Materials and Methods CTP was performed in 160 cases suspected with liver disease, who were randomly divided into 2 groups and scanned with 3mm collimation (n=98) and 5mm (n=62), respectively. The contrast dose was 2ml/kg weight with the injection rate of 3ml/s. The delay time was within 45~50s. Both MIP and MPVR were used to reconstruct the portal veins in all patients. The images were evaluated and scored based on the grade of portal vein, CT value difference between the portal vein and the liver parenchyma, the definition of blood vessel edge, the grade of hepatic vein and the display of IVC by two radiologists.Results The average combined score of MIP images in 3mm and 5mm groups was 10.8 and 8.9, while MPVR was 12.5 and 10.7, respectively. For visualizing the portal vein and hepatic vein and the clearness of blood vessel margin, 3mm collimation group and MPVR were superior to 5mm group and MIP.Conclusion In spiral CT portography, it is of optimal choice to use 3mm collimation and MPVR reconstructive method.