The Discussion of Scanning Techniques Using Dual-Source CT with High-Pitch Mode for Volume Rendering Reconstruction of Pulmonary Vessels

杨凤娇,张艳,吴建伟,汪栋,李如帅
DOI: https://doi.org/10.13437/j.cnki.jcr.2017.03.003
2017-01-01
Abstract:Objective To discuss the scanning techniques using dual-source CT with high-pitch mode for volume rendering(VR) reconstruction of pulmonary vessels.Methods The study included 85 patients who had contrast-enhanced dual-source CT scan with high-pitch mode for volume rendering reconstruction of pulmonary vessels.The bolus tracking technique was applied with the ROI placed within the brachiocephalic veins.All subjects were divided randomly into three groups according to delay time:4 seconds-delay group (group 1),5 seconds-delay group (group 2) and 6 seconds-delay group(group 3).CT value of pulmonary arteries,left lower pulmonary arteries and left upper pulmonary veins were measured.Among the three groups,the signal-to-noise ratio(SNR),contrast-to-noise ratio(CNR) and the differences of CT value between left lower pulmonary arteries and left upper pulmonary veins were calculated and compared.VR reconstruction was done and their image-grading criteria was as follows:excellent-above subsegmental pulmonary artery represented as red,pulmonary veins in blue,good-above segmental pulmonary artery presented rad,subsegmental artery and pulmonary veins in blue,poor-impossible to distinguish the colors between arteries and veins.The differences of VR image grades among the three groups were also compared.Results The SNR and CNR had significant differences between group 1 and group 2,group 1 and group 3.Among the three groups,Ridit analysis of VR images grading showed that the average Ridit values were 0.609,0.391 and 0.489 respectively,in which the VR images of group 2 were the best with minimal Ridit value,the percentage of excellent plus good images was 75.9%,96.2% and 76.7% respectively(x2 =4.916,P =0.086).The difference of CT values between arteries and veins of the grade excellent,good and poor images were (228.63 ± 83.551)HU,(155.38 ±38.83)HU and (88.07 ±54.53)HU respectively.Conclusion Dual-source CT can enhance the pulmonary angiography using a high-pitch scan and set up the monitoring site of the brachiocephalic veins.The quality of the pulmonary veins in the VR images can be improved by a delay of 5 seconds.
What problem does this paper attempt to address?