Establishment and value of the optimal technical protocol of whole liver perfusion with 128-CT

TANG Min,WU Dong,ZENG Meng-su,CHEN Gang,NI Xiao-yan,HUANG Hui
DOI: https://doi.org/10.3969/j.issn.1672-8467.2013.03.014
2013-01-01
Abstract:Objective To optimize protocol of whole liver CT perfusion(CTP) with 128-CT.Methods In 90 patients received whole liver CTP,pair analysis was used to compare image quality under different tube electric voltage(100 kVp vs.120 kVp),different contrastation of contrast media(300 mgI/mL vs.370 mgI/mL),different injection rate of contrast media(5 mL/s vs.8 mL/s) and with or without motion correction and filtration.Liver artifact,tissue noise,time density(TDC) peak time and peak value of hepatic artery,portal vein and spleen were evaluated.Repeatability test for data collected from the same observer was also done.Results Two-way analysis of variance(ANOVA) indicated artifact and tissue noise had no correlation with different tube electric voltage(P>0.05),but showed statistical difference in different lobe(P<0.001).When the concentrations of contrast media were 370 mgI/mL and 300 mgI/mL,peak value were(391.33±33.57) Hu and(288.67±32.26) Hu for hepatic artery,(124.13±17.66) Hu and(78.67±21.91) Hu for portal vein,(111.33±17.16) Hu and(59.2±17.78) Hu for spleen,and the differences were statistically significant(P <0.001).When the injection rates of contrast media were 5 mL/s and 8 mL/s,peak time were(13.33±1.84) s and(9.07±1.16)s for hepatic artery,(30.95±5.87) s and(19.2±2.46) s for portal vein,(22.47±4.67) s and(14.87±2.13) s for spleen,and the differences were statistically significant(P<0.0001).Artifact scores before and after the application of motion correction and filtration were 1.9±0.3 and 0.6±0.3,and the difference was statistically significant(P<0.001).There was no significant difference in CTP parameter sabtained at different time points by the same observer.Conclusions 128-CT could be used for whole liver CTP.Tube electric voltage of 100 kVp provides good images,when combined with 8 mL/s injection rate,370 mgI/mL contrast media concentration,and with motion correction and filtration,which could be an optimal technical protocol.The repeatability is also excellent in the same observer.
What problem does this paper attempt to address?