Amperometric Micro-Biosensor Based on Cholesterol Determination

管文军,胡贵权,李光,廖静敏,陈裕泉
DOI: https://doi.org/10.3969/j.issn.1004-1699.2003.03.015
2003-01-01
Abstract:Printed graphite electrode based micro biosensor containing cholesterol oxidase, cholesterol esterase and peroxidase was developed. It was used for whole blood total cholesterol determination. Enzymes in reactive system were immobilized on the surface of graphite electrode through trehalose embedding. The biosensor's linear range is 100~400 mg/dl, sample volume is 2 microlitre and detecting time is 180 s. The problem of inactivation of bioactive substances has been solved and stable storage time of the biosensor is more than six months.
What problem does this paper attempt to address?