A Whole-Blood L-Lactic Acid Biosensor

吴维明,童基均,管文军,李光,潘敏,陈裕泉
DOI: https://doi.org/10.3969/j.issn.1004-1699.2003.04.025
2003-01-01
Abstract:A whole-blood L-lactic acid biosensor was developed for measuring the L-lactic acid in whole blood rapidly and conveniently. The sensor was made of a strip with two screen-printed electrodes and a reaction layer consisting of L-lactate oxidase and ferricyanide and a reaction chamber formed by plastic. A voltage (300mV) was applied between the two electrodes of the sensor after a blood sample(2ul) was dipped in the reaction chamber. The current was measured to obtain the concentration of the L-lactic acid after 50s. The linear range of the sensor was up to 15mM. The activity of the sensor remained high up to 120 days. L-lactic acid concentration measured by this method agreed with the result obtained by stand instrument VITROS250 SYSTEM(r=0.9809).The properties of the biosensor indicated that the biosensor is feasible for mass production.
What problem does this paper attempt to address?