A Novel Oxidase Electrode for Determination of Glucose Goupled with Capillary Electrophoresis

陈刚,叶建农,张剑霞,孔继烈
DOI: https://doi.org/10.3969/j.issn.0427-7104.2002.04.008
2002-01-01
Abstract:Ferroceneethylamine and chitosan were modified on a platinum eletrode by dipping method. And then,the modified electrode was sunk is glucose oxidase solution to immobilize the enzyme into the chitosan membrane. A novel glucose oxidase biosensor could be fabricated after it was finally treated with cupling reagent,glutaraldehyde. The glucose V A curve on the target biosensor has been measured. Using amperometric determination method with potential keeping constant,the current response of the sensor is found to be linear with the glucose concentration in the range of 0 002-4 0 mmol/L with fast responese. Since 1 ferroceneehtylamine and glucose oxidase are,through the action of warp and bond,immobilized in the chitosan membrane,being not easy to be lost from electrode surface,the electrode owns high stability. It has been coupled with capillary electrophoresis to determine the glucose in human serum with satisfactory assay results.