A Glucose Biosensor Based on Chitosan–glucose Oxidase–gold Nanoparticles Biocomposite Formed by One-Step Electrodeposition

XL Luo,JJ Xu,Y Du,HY Chen
DOI: https://doi.org/10.1016/j.ab.2004.07.005
IF: 2.9
2004-01-01
Analytical Biochemistry
Abstract:An amperometric biosensor for the quantitative measurement of glucose is reported. The biosensor is based on a biocomposite that is homogeneous and easily prepared. This biocomposite is made of chitosan hydrogel, glucose oxidase, and gold nanoparticles by a direct and facile electrochemical deposition method under enzyme-friendly conditions. The resulting biocomposite provided a shelter for the enzyme to retain its bioactivity at considerably extreme conditions, and the decorated gold nanoparticles in the biocomposite offer excellent affinity to enzyme. The biosensor exhibited a rapid response (within 7s) and a linear calibration range from 5.0μM to 2.4mM with a detection limit of 2.7μM for the detection of glucose. The combination of gold nanoparticles affinity and the promising feature of the biocomposite with the onestep nonmanual technique favor the sensitive determination of glucose with improved analytical capabilities.
What problem does this paper attempt to address?