Quality assurance for multileaf collimator with radiographic film exposed by slit beam

马金利,蒋国梁,傅小龙,廖源,吴开良,周莉钧
DOI: https://doi.org/10.3760/j.issn:1004-4221.2004.02.013
2004-01-01
Abstract:Objective To evaluate the role of Kodak X-OMAT-V film exposed by slit beam in the check of various leaf positions of multileaf collimator(MLC), and to check the status of Varian 26 leaf pairs MLC in the Department of Radiation Oncology in Shanghai Cancer Hospital affiliated to Fudan University. Methods At first, some position errors of different sizes were produced for different leaves so as to determine the minimal leaf position error that could be seen on film. Then, exposure conditions including the exposure dose and source to film distance were changed to find the optimal one. Finally, a Kodak X-OMAT-V film was exposed with a leaf sequence file which was designated randomly by a physicist with leaf position errors of different sizes. After the film was developed, two doctors and two physicists were invited to observe, on blind basis, in order to determine the sensitivity and specificity of the film in the check of leaf positions. Ultimately, leaf positions of the Varian 26 leaf pairs MLC were checked, in which way, the leaf motor status and the carriage stability were checked indirectly. Results Leaf position errors no less than 0.2?mm could be found using Kodak X-OMAT-V film under the following conditions: source to film distance 100?cm, exposure dose 25?MU, which had been considered as the optimal exposure conditions. The sensitivity and specificity of this method were 73.4% and 96.4%. Any MLC leaf position errors more than 0.2?mm could not be detected. Thus, it was deemed that all leaf motors of the Varian 26 leaf pairs MLC were well in gear and the carriages were stable. Conclusions MLC leaf position errors can be detected by Kodak X-OMAT-V film exposed by slit beam with high accuracy, but the ability to find leaf position errors with the naked eye may vary from person to person. It is proposed that the Kodak X-OMAT-V film exposed by slit beam be used to check the MLC leaf positions, i.e. the leaf motor status and carriage stability, at regular intervals under the optimal conditions determined in this study.
What problem does this paper attempt to address?