Comparison of the Positioning Accuracy in MLC Quality Assurance using Various Detectors

WU Hao,REN Min,HUANG Fei-zeng,BAO Shang-lian
DOI: https://doi.org/10.3969/j.issn.1005-202x.2010.04.003
2010-01-01
Abstract:Objects: To compare the positioning accuracy among GAFCHROMIC EBT film,Portal Vision and ionchamber 2D array in the MLC quality assurance.Method: Measuring the edge response function to each leaf for Varian IX Linac MLC,respectively with GAFCHROMIC EBT film,Portal Vision and ionchamber 2D array(MatriXX and Seven 29 each belongs to IBA and PTW),compare,and evaluate their positioning accuracy of each detector.Result: The position accuracy of each detector all reached to 0.1mm accuracy,the sensitivity of Portal Vision is the best.Conclusion: All detectors we used could meet the clinical needs for quality control for MLC positionning.