Micropressure Transducer with High Over-Load Capacity

YIN Hao,Yang Heng,WANG Wen-xiang,SHEN Shao-qun,BAO Min-hang
DOI: https://doi.org/10.3969/j.issn.1000-9787.2001.04.015
2001-01-01
Abstract:In the process of improving the sensitivity of a pressure transducer, one always has the problems such as the deterioration of linearity and stability, insufficiency of over-load capacity and the sticking of the back islands to the glass in the electrostatic bonding process. Two structure improvements have been made to overcome these difficulties: the structure for stress unification on diaphragm and the two-leveled island for over-range protection.Both finite element analysis and analytical analysis are used in the design of the high sensitivity pressure transducer with novel two-leveled islands and beam-diaphragm structure. A high sensitivity pressure transducer based on the structure has been developed and tested.The measument range is 0.3?kPa,the accuracy of the devices is 0.4% and the overload capacity is larger than 140 times of the full operation range.