Determination of Anastrozole in Human Plasma by HPLC-MS/MS

PAN Xia-yun,JING Xian,CHENG Gang,ZENG Ling,CHEN Yao,WANG Yi-cheng,ZHOU Gan,TAN Zhi-rong
DOI: https://doi.org/10.3969/j.issn.1672-2981.2011.06.011
2011-01-01
Abstract:Objective To establish an HPLC-MS/MS method for the determination of anastrozole in human plasma.Methods Plasma sample was extracted by MTBE and letrozol was chosen as the internal standard.The separation was performed on a Thermo Hypurity C18(2.1 mm×150 mm,5 μm) column.The mobile phase was acetonitrile-0.1% formic acid(60∶40,v/v) at 0.20 mL·min-1.The analytes were detected by the electronic-spray ionization(ESI)-multiple reaction monitoring(MRM) mode.MS detection parameters:the source voltage was 3.6 kV and the source temperature was 100 ℃.Anastrozole and letrozol ion selective channels were: m/z294.69→225.41,286.25→217.60.Results The linear range for anastrozole was 0.214-214 ng·mL-1 and the minimum detectable concentration was 0.214 ng·mL-1.The mean relative recovery was 80%-120% and the relative standard deviation of intra-day and inter-day was less than 15%.Conclusion This method is simple and sensitive,which can be used to determinate the concentration of anastrozole in human plasma.
What problem does this paper attempt to address?