Determination Of Diethylstilbestrol In Human Plasma With Measurement Uncertainty Estimation By Liquid Chromatography-Tandem Mass Spectrometry

Yi Liu,Na Li,Liping Ma,Lin Huang,Shan Jing,Libo Zhao,Wanyu Feng
DOI: https://doi.org/10.1080/10826076.2012.745140
2014-01-01
Abstract:A rapid and sensitive liquid chromatography/tandem mass spectrometry (LC-MS/MS) using an electrospray ionization (ESI) method was developed and validated to determine the concentration of diethylstilbestrol in human plasma. Samples were extracted with methyl tert-butyl ether after alkalified by sodium tetraborate solution, using 3,5-dihydroxy-4-isopropyl toluylene as the internal standard (IS). Separation was performed on a Restek C-18 column (2.1mmx150mm, 5.0 mu m) guarded by a Hypersil Gold pre-column (2.1mmx10mm, 3 mu m). The mobile phase consisted of acetonitrile-water (85:15, containing 0.2mmol/L ammonium acetate) at a flow rate of 0.3mL/min. Detection was performed using negative multiple reaction monitoring (MRM) mode. The mass transitions monitored were [M-H](-)m/z267.2237.2 and [M-H](-)m/z253.2211.0 for diethylstilbestrol and IS, respectively. The method was fully validated using total error theory, which is based on -expectation tolerance intervals and include trueness and intermediate precision. The method was found to be accurate over a concentration range of 0.1-50ng/mL. The main advantages of this method were the high sensitivity (a lower limit of quantification, LLOQ, of 0.1ng/mL) and the short analysis time (total run time 5min per sample). The measurement uncertainty based on -expectation tolerance intervals was assessed at each concentration level of the validation standards. The method is specific, sensitive, and accuracy, and it is suitable for plasma concentration determination of diethylstilbestrol.
What problem does this paper attempt to address?