Establishment of a liquid chromatography-tandem mass spectrometry for the determination of plasma 3-methoxytyramine

Jiaqian QIN,Yingfei PENG,Fangjun CHEN,Wenqi SHAO,Jiong WU,Chunyan ZHANG,Beili WANG,Wei GUO,Baishen PAN
DOI: https://doi.org/10.3969/j.issn.1673-8640.2017.011.024
IF: 1.865
2017-01-01
Laboratory Medicine
Abstract:Objective To establish a liquid chromatography-tandem mass spectrometry(LC-MS/MS) for plasma 3-methoxytyramine(3-MT) determination. Methods The separation condition of LC-MS/MS(chromatographic column,column temperature and pH value)was improved. The linear range,recovery rate,precision,lower limit of quantitation and stability of LC-MS/MS were evaluated. Results 3-MT was separated by BEH HILIC chromatographic column using 3-MT-d4 as internal standard. The mobile phase consisted of 100 mmol/L ammonium formate aqueous solution(pH value was 3)and acetonitrile. The column temperature was 35 ℃. The linear range of 3-MT determination were 5-1000 pg/mL. The limit of quantitation was 5 pg/mL. The inter-day and between-run coefficients of variation(CV) were <6% and <7%. The recovery rate was 97.1%-109.3%. As 3-MT was unstable in room temperature,it should be transported under 4 ℃.?Conclusions A reliable LC-MS/MS for the determination of 3-MT has been established,with sensitivity and accuracy.
What problem does this paper attempt to address?